PENGUKURAN KETULENAN EMAS MENGGUNAKAN KAEDAH UJIAN TANPA MUSNAH: SUATU ULASAN PERBANDINGAN

Authors

  • FAZRUL MOHD NOR Seksyen Termoplastik, Institut Metrologi Kebangsaan Malaysia, SIRIM Berhad, Malaysia
  • ABD RAHMAN TAMURI Fakulti Sains, Universiti Teknologi Malaysia, Malaysia
  • ABD KHAMIM ISMAIL Faculty of Science, Universiti Teknologi Malaysia, Malaysia
  • MOHD ERFY ISMAIL Faculty of Technical dan Vocational Education, Universiti Tun Hussein Onn Malaysia,, Malaysia
  • MOHD HASRIL AMIRUDDIN Faculty of Technical dan Vocational Education, Universiti Tun Hussein Onn Malaysia, Malaysia

DOI:

https://doi.org/10.47252/teniat.v9i2.745

Keywords:

Emas Palsu, Jongkong Emas, Ketumpatan, Tungsten, Ujian Tanpa Musnah, Fake Gold, Gold Bar, Density, Non-Destructive Test

Abstract

Abstrak
Penentuan ketulenan jongkong emas dengan kaedah ujian tanpa musnah adalah perkara yang paling mencabar dalam industri emas terutama jika jongkong emas dipalsukan dengan cara meletakkan tungsten di dalamnya. Instrumen yang sedia ada mempunyai keterbatasan untuk menentukan ketulenan jongkong emas dan sukar untuk mengesan tungsten di dalamnya kerana tungsten mempunyai ketumpatan yang hampir sama dengan emas. Kajian ini mengemukakan keterbatasan instrumen kaedah ujian tanpa musnah seperti densimeter, alat timbangan, sinar-x pendarfluor dan ultrasonik. Dalam kajian ini menemui kesukaran untuk mengenal pasti emas palsu menggunakan instrumen kaedah ujian tanpa musnah seperti ultrasonik dan x-ray fluorescence (XRF) kerana instrumen ini mempunyai batasan. Setakat ini tiada lagi kajian yang berkaitan pemalsuan jongkong emas kerana harganya yang mahal untuk dijadikan sebagai sampel kajian. Masalah jongkong emas palsu dapat diselesaikan jika ketepatan instrumen yang digunakan untuk mengukur ketumpatan dapat ditingkatkan lagi sehingga 0.01 g/ml. Ini kerana perbezaan antara ketumpatan emas dan tungsten adalah 0.05 g/ml. Pemalsuan emas dengan mencampurkan tungsten akan dapat dikesan jika ada instrumen yang mempunyai ketepatan sekurang-kurangnya 0.01 g/ml. Manakala instrumen lain seperti XRF digunakan untuk mengesahkan ketulenan emas di bahagian permukaan sahaja. Oleh itu, bagi pengujian yang lebih menyeluruh masalah jongkong emas palsu ini perlu diselesaikan dengan menggabungkan kedua-dua kaedah ini.

 

Abstract

Determination of the gold bar purity by non-destructive method remains one of the most challenging in gold industry especially when the tungsten is inside the gold bar. The existing instruments have limitations in their accuracy to determine the purity of gold when tungsten also inside the gold bar since both tungsten and gold have almost similar densities. This study presents the limitations of the non-destructive test method instruments such as densimeter, weighing balance, x-ray fluorescence (XRF) and ultrasonic in determining the purity of gold. In this study found the difficulty in identifying counterfeit gold using non-destructive test method instruments such as ultrasonic and XRF due to some limitations. So far, there are no studies related to the counterfeiting of gold bars due to its expensive price to be used as a study sample. The problem of counterfeit gold bars can be solved if the accuracy of the instrument used to measure the density can be further increased up to 0.01 g/ml. This is because the difference between the density of gold and tungsten is 0.05 g/ml. Gold counterfeiting by mixing tungsten will be detectable if there is an instrument that has an accuracy of at least 0.01 g/ml. While other instruments such as XRF will be used to verify the purity of gold on the surface only. Therefore, for more thorough testing the problem of fake gold ingots needs to be solved by combining these two methods.

Downloads

Published

2021-09-29

How to Cite

MOHD NOR, F. ., TAMURI, A. R. ., ISMAIL, A. K. ., ISMAIL, M. E. ., & AMIRUDDIN, M. H. . (2021). PENGUKURAN KETULENAN EMAS MENGGUNAKAN KAEDAH UJIAN TANPA MUSNAH: SUATU ULASAN PERBANDINGAN. International Journal of Creative Future and Heritage (TENIAT), 9(2), 91–106. https://doi.org/10.47252/teniat.v9i2.745